The high-quality sensors enable reliable measurement of all dissolved oxygen concentrations from the trace level to the saturation level. Both versions have been especially developed for hygienic as well as demanding applications and represent different measuring ranges. Common field devices and process control systems can be connected via an integrated analog and digital interface. The connection to the intelligent, bus-compatible JUMO digiLine system is made via Plug and Play.
Use even in harsh and demanding environments
Maximum plant safety and process reliability
Fast conversion of existing systems
Use in a wide variety of measurement applications
Process reliability even under demanding operating conditions
Application also in direct contact with food and pharmaceutical substances
general technical data | |
---|---|
Sensor material | Stainless steel 1.4435 |
Measuring principle | Optical |
Dimensions | 12 mm diameter, 120 mm shaft length; other lengths upon request |
Connection | VP plug head Pg 13.5 |
Measuring range | General: 1 ppb (µg/L) to 45 ppm (mg/L)* 0 to 500 % SAT* 0 to 100 % O2* 0 to 515 hPa* *at 20 °C, 1013 hPa in air JUMO digiLine H10: standard version* 0 to 22.5 ppm (mg/L) 0 to 248 % SAT 0 to 50 % O2 *at 20°C, 1013 hPa in air JUMO digiLine H20: standard version* 0 to 2000 ppb (µg/L) 0 to 5 % O2; 0 to 50 hPa *at 20 °C, 1013 hPa in air |
Temperature compensation | Automatic |
Process temperature | 0 to 50 °C optimally; -10 to +90 °C max. |
Process pressure | At 20 °C: -1 to +12 bar |
CIP | 2 % NaOH, 80 °C |
SIP | Up to 140 °C, 1.5 atm |
Outputs | Digital; 4 to 20 mA |
Serial interface | RS485 with Modbus RTU protocol |
Cross sensitivity | Chlorine gas, organic solvents such as acetone, toluene, chloroform, chloroacetic acid |
Special features | JUMO digiLine connection system |