The capacitive level switch identifies the level of liquids and solids. It can also be used in pressurized tanks or pipes. Typically, the level switch is used in industrial environments for applications with requirements for overflow and dry-run protection or for media detection (such as sludge in separation processes).
general technical data | |
---|---|
Switching capacity | Up to DC 36 V and 200 mA |
Medium temperature | -40 to +200 °C |
Repeatability and hysteresis | +/- 1 mm in each case |
Output signals | PNP, NPN, push-pull, IO-Link |
Parts in contact with medium | PEEK |
Additional equipment | unique auto-calibration function as standard, short-circuit and reverse polarity protection, 360° status display according to NE107, IO-Link or customer-specific |
Protection type | IP67/IP69 |
Approvals | ATEX, IEC Ex [Ex i], DNV, WHG, EHEDG, 3A, UL |
Declaration of manufacturer | Railway technology test according to DIN EN 50155, CE, RoHS |
679.39 KB, English